RRR - Bulk material

Figure
Detailed photograph of the experimental platform, indicating instrumentation and showing the six samples mounted onto the platform.

Insert placed in a vacuum vessel. Samples are cooled via a phase separator continuously fed to provide evaporative cooling. The samples reach a minimum temperature of 6 K, with a thermal shield enclosing the experimental platform. Up to six samples can be measured simultaneously. The maximum supply current is 5 A.

The samples need to have the following characteristics:

  1. Maximum curvature radius of 1 m;
  2. Length comprised between 105 mm and 120 mm;
  3. Width comprised between 3 mm and 6 mm;
  4. Thickness comprised between 1 mm and 2.5 mm with parallel surfaces to ensure uniform thermal contact to the experimental platform.

 

L’insert est placé dans la chambre à vide. Les échantillons sont refroidis par l’intermédiaire d’un séparateur de phase alimenté en permanence pour assurer une constante circulation des vapeurs froides.  Grâce à la mise en place d’un écran thermique autour du système, les échantillons peuvent atteindre une température de 6 K. Il est possible de mesurer jusqu’à 6 échantillons en même temps. Le courant électrique maximal est de 5 A.

Les échantillons doivent avoir les caractéristiques suivantes :

  1. Rayon de courbure maximal de 1m ;
  2. Longueur comprise entre 105 et 120 mm ;
  3. Largeur comprise entre 3 et 6 mm ;
  4. Épaisseur comprise entre 1 et 2,5 mm avec des surfaces parallèles assurant un contact thermique uniforme sur le support expérimental.